Evaluation of a Micro CT scan system based on digital image correlation profile
11 Halaman
Penulis
ISSN
2685-8673 (Online)
Penerbit
Universitas Aisyiyah Yogyakarta, Indonesia
Diterbitkan pada
01/11/2022
Bahasa
English
Kata Kunci
Abstrak
The centreof rotation position is a factor that needs to be controlled to avoid artefactin CT scan. Amethod to evaluate the performance of CT system is needed. The method is developed using x-ray micro-CT system and evaluated based on Digital Image Correlation(DIC). The method of this research is experiment. The multiple radiographs are obtained from scanning object by simulation and experiment.The results show that the ideal DIC profile value of radiograph image from simulationis a constant straight line for every size variation of the object. However, from the experiment, the result show deviation toward –y for every size variation of the object that indicate the shift of COR position in micro-CT system. The result also similar using centreof mass profile value as comparison.