Transient conduction disturbances acutely after pulsed-field cavotricuspid isthmus ablation: a case report

5 Halaman

Penulis

, , ,

ISSN

2514-2119 (ONLINE)

Penerbit

Oxford University Press , United Kingdom

Diterbitkan pada

02/08/2023

Bahasa

English

Kata Kunci

, , , , ,

Abstrak

Background Cavotricuspid isthmus pulsed-field ablation has been recently described to be safely performed despite initial reports on coronary arterial spasm while conduction disturbances as a complication of cavotricuspid isthmus ablation are rare and have been reported exclusively for radiofrequency catheter ablation. Case summary A 64-year-old female patient with mechanical prosthetic valves underwent atrial fibrillation ablation using the pentaspline pulsed-field ablation catheter. At the end of the uneventful pulmonary vein isolation, an atrial tachycardia depended to the cavotricuspid isthmus occurred. A single pulsed-field application at the cavotricuspid isthmus resulted in right bundle branch block combined with posterior fascicular hemiblock and PR prolongation that resolved spontaneously within 12 h. Discussion This is the first report of transient conduction disturbances as a complication of cavotricuspid isthmus pulsed-field ablation. Although the underlying mechanism, either single or miscellaneous, was not verified, this case highlights that caution should be taken when the pentaspline pulsed-field ablation catheter is used for cavotricuspid isthmus ablation.