Critical Impact of Storage Conditions on Kidney Biomarkers Dampak Kritis Kondisi Penyimpanan pada Biomarker Ginjal

9 Halaman

Penulis

, , ,

ISSN

3063-8186

Penerbit

Universitas Muhammadiyah Sidoarjo

Diterbitkan pada

19/06/2024

Bahasa

Indonesia

Kata Kunci

, , , ,

Abstrak

This study investigated the impact of storage time (0, 4, 8, and 24 hours) at 2-8°C on urea and creatinine levels in serum samples from healthy young adults. Using a laboratory experimental design, 32 serum samples were analyzed with enzymatic and Jaffe-kinetic methods on a Microlab 300 photometer. Results from One Way ANOVA indicated significant effects on urea (p=0.007) and creatinine levels (p=0.009) across different storage times, with Post-Hoc tests revealing significant differences between the 0-hour and 24-hour storage groups for both biomarkers. These findings highlight the importance of standardized storage protocols in clinical laboratories to ensure accurate kidney function assessment and improve diagnostic reliability.