Analysis of Viability as Readout of Lymphocyte Transformation Test in Drug Hypersensitivity Diagnostics

12 Halaman

Penulis

, , ,

ISSN

2313-5786

Penerbit

MDPI

Diterbitkan pada

09/01/2025

Bahasa

English

Kata Kunci

, , ,

Abstrak

In vitro tests of cellular activity form part of the diagnostic algorithm of drug hypersensitivity reactions. Because of the wide range of pharmacological mechanisms, clinical symptoms, genetic components, and laboratory tests involved, it is important to know how a particular test performs in the diagnostic procedure. We carried out a detailed retrospective analysis of more than 6000 measurements of numerous drug compounds tested in 738 serum samples over the past 6 years. Our cell viability-based lymphocyte transformation had a coefficient of variation of 10% and showed similar performance over the whole range of tested ages. With an adequate number of parallel measurements, the test can identify modest increases in stimulation indices with high confidence. Similar percentages of analytically positive responses (11.4%, 13.5%, and 9.7%) were observed for the three most frequently tested drug groups, namely, antibiotics, non-steroid antiinflammatory agents, and anesthetics. These results confirm that cell viability tests are suitable alternatives for proliferation assays in drug allergy testing.

Kata Mereka

Testimoni Pengguna M3

Lebih Banyak