High Precision Weighing and Uncertainty Assessment in Reagent Preparation
8 Halaman
Penulis
ISSN
2753-8826
Penerbit
EWA Publishing
Diterbitkan pada
08/03/2024
Bahasa
English
Kata Kunci
Abstrak
The volume of the reagent in one tube was difficult to be measured in reagent preparation, as the volume of each portion was extremely small. Formerly, the measurement errors of direct weighing method were large. Therefore, the method of twice minusing method was used to weighing in order to obtain the reagent preparation result more accurately. The sources of uncertainty of this method were analyzed and the uncertainty was evaluated, the uncertainty of twice minusing method and direct weighing method was compared. The results showed that the expanded uncertainty of the reagents in one of the eight-connected tubes was 2.266×10-4 g (k = 2) when using twice minusing method, and the expanded uncertainty was 1.704×10-3 g (k = 2) when using direct weighing method, so the results of twice minusing method were more reliable.