Determination of Complete Blood Count Reference Values of Mindray BC-760 Hematology Analyzer

11 Halaman

Penulis

, , , , , , , , ,

ISSN

2621-1122 (ONLINE)

Penerbit

the Faculty of Medicine UPN Veteran Jakarta

Diterbitkan pada

15/06/2024

Bahasa

English

Kata Kunci

, ,

Abstrak

Hematological assessment serves as a standard examination in supporting diagnostics in clinical practice. The advanced hematology analyzer instrument namely Mindray BC 760 has recently introduced several new hematological parameters which previously unavailable. This study set out to establish the value of the reference interval of those new parameters on the Mindray BC 760 device in the Indonesian adult population. A total of 352 subjects who underwent medical check-ups at pathology clinics laboratory in the hospital were enrolled. All participants comprised both females and males aged > 17 years and were confirmed to be healthy through examination. Hematological was assessed using Mindray BC 760 device. It is recommended to adjust of reference interval value according to gender for several novel hematological parameters assessed through the Mindray BC 760 instrument. Significant gender-based differences (p < 0.05) were found in multiple new parameters such as Hb, Ht, RBC, MCH, MCHC, RHE, RDW-CV, RET#, MON#, EOS#, NEU%, MON%, EOS%, PLT_I, PLT_H, PLT_O, PDW, PCT, and PLCC, suggesting the need for gender-specific reference intervals. Conversely, certain parameters showed no significant gender differences (p ≥ 0.05) including MCV, RDW-SD, RET%, LFR, MFR, HFR, NRBC#, NRBC%, WBC, NEU#, LYM#, BAS#, IMG#, LYM%, BAS%, IMG%, MPV, PLCR, and IPF.

Kata Mereka

Testimoni Pengguna M3

Lebih Banyak