The Comparison of Surface Dose in Vivo Dosimetry Using 3D-CRT and IMRT Technique on Breast Cancer Case

10 Halaman

Penulis

, , , ,

ISSN

2621-1122 (ONLINE)

Penerbit

the Faculty of Medicine UPN Veteran Jakarta

Diterbitkan pada

15/06/2024

Bahasa

English

Kata Kunci

, , ,

Abstrak

Verification of surface radiation dose in breast cancer cases using the in vivo dosimetry method with TLD-100 at the Radiation Oncology Installation of Andalas University Hospital has been conducted. The aim of this study is to verify the adequacy of the surface radiation dose calculated by the Treatment Planning System (TPS) using 3DCRT and IMRT technique with the dose measured by the TLD-100, referring to the report of the American Association of Physicists in Medicine Task Group No.219 (AAPM-TG No.219). This research process began with annealing the TLD 100, followed by scanning the TLD-100 on the surface of the slab phantom using a CT Simulator. In addition, the TLD-100 was calibrated with different radiation doses (0; 20; 50; 80; 100; 150; 200; 250; 370) cGy. Calculating the TPS surface dose by Patient Specific Quality Assurance (PSQA) and comparing it with the dose measured by the TLD-100 are done to verify the surface radiation dose. The result is radiation techniques such as IMRT are more conformal than 3D-CRT because they precisely control dose intensity, making the TLD dose in IMRT closer to that of TPS. Both techniques have discrepancy between TLD and TPS measurement is below the AAPM-TG 219 tolerance limit, with an average discrepancy of 2,84% for IMRT and 3,15% for 3D-CRT.