Validity test of POCT (Point of Care Testing) method on blood glucose examination using whole blood samples, serum, and EDTA plasma
8 Halaman
Penulis
ISSN
2829-8985
Penerbit
Lighthouse Publishing Company
Diterbitkan pada
31/05/2022
Bahasa
English
DOI
-Kata Kunci
Abstrak
Medical laboratory services play a crucial role in disease diagnosis and monitoring patient recovery. Point-of-Care Testing (POCT) methods are increasingly used in clinical laboratories, not only for screening but also for diagnosing conditions such as diabetes mellitus. This study aims to assess the validity of the POCT method for blood glucose testing using different sample types: whole blood, serum, and plasma EDTA. The research employed a descriptive-analytical approach to evaluate the performance of the POCT method across these three types of samples. Precision was tested with normal and pathological serum samples, yielding Coefficient of Variation (CV) percentages of 2.02% and 2.27%, respectively. Accuracy was also evaluated with normal and pathological serum samples, showing Total Error (TE) percentages of 8.54% and 6.03%. Linearity was confirmed for serum and plasma EDTA samples with an R² value of 0.998. Despite these positive results, sigma values fell into the unacceptable range. This indicates that while the POCT tool shows valid performance for blood glucose examination, deviations in test results are influenced by pre-analytical errors such as sampling and sample processing. Consequently, the total error exceeded the acceptable threshold. Overall, the POCT tool is applicable for all types of samples, but attention to pre-analytical procedures is essential to ensure accuracy.
